Main Article Content
Sequential injection spectrophotometric determination of V(V) in environmental polluted waters
Abstract
Linear calibration plots were obtained for V(V) concentrations between 0.50 and 5.0 mg..-1. The developed methodology exhibits a good precision, with relative standard deviation (rsd) < 2.0% (n=15) and the detection limit was 0.39 mg..-1. The presented SIA methodology was applied to the determination of V(V) in 10 water samples and in a wastewater
reference certified sample. The determination of V(V) by the developed automatic procedure involved the consumption of 1.9 mg of 4AAP and
2.9 mg of DP and the production of 2.35 m. of effluent.