Main Article Content

Glass Electrode Calibration for Use in the Voltammetric Determination of Stability Constants under Extreme Acidic Conditions


C Billing
I Cukrowski

Abstract

A glass electrode (GE) can be successfully employed to measure pH in the study of metal-ligand equilibria by voltammetry at extremely low pH (between 0 and 2); two consecutive strong acid–strong base titrations involving different base concentrations (recommended to avoid corrosion of the GE in very basic solutions) are best suited to establish the response parameters of a GE. A novel approach of using a combined linear and binomial GE calibration was developed; this procedure allows measurements between pH 0 and 2 with uncertainty better than ±0.01 pH unit. From an extensive error analysis, it has been established that the uncertainties of about ±0.5 mV in the response slope and ±1.3 mV in E°’ might result in an absolute error in pH of about 0.02 which should not generate errors larger than 0.3% in optimized stability constants (as logK values) determined by voltammetry at extremely low pH. A test of GE suitability for the study of metal complexes by voltammetry is also proposed; it should be implemented only for suspect electrodes that show response parameters outside the limits recommended in this work.

Keywords: Glass electrode calibration, glass electrode performance, use of GE in highly acidic media, metal-ligand equilibria studies, potentiometry, voltammetry