Outcomes of the Icare Rebound Tonometer versus the Gold Standard Tonometer in Measuring Intraocular Pressure among Mzuzu University Students, Malawi


François Régis Twiringirimana
Clémentine Kanazayire
Okyere Bright Vandyke
Andrew Kanzunguze
Reverien Rutayisire
Grace Ogbona

Abstract

Introduction
The Goldmann Applanation Tonometer (GAT) is accepted as the gold standard tool to measure Intraocular Pressure (IOP). Besides, there is an emerging Icare Rebound Tonometer (RBT) which can be an alternative in terms of measuring IOP.
Objective
This research aimed to analyze the agreement, and repeatability of intraocular pressure (IOP) measurements and compare the outcomes of RBT and GAT among Mzuzu University Students in Malawi.
Methods
For this cross-sectional study, 107 participants aged 18 to 29 years underwent ophthalmic examination with three IOP measurements taken by two examiners using the RBT and the GAT.
Results
The RBT and GAT were found to have a strong correlation (r=0.809, p>0.001), with an intra-class correlation of 0.787 (p>0.001). The mean difference of IOP measured by the two machines was 1.99±1.05mmHg, which was statistically significant. However, according to Bland-Altman analysis, there was no agreement between the results given by the two machines, with a bias of -1.979 (CI -2.28, -1.68, p > 0.001). In terms of repeatability, GAT was found to be better than RBT, with a CR of 7.512 and 10.418 respectively.
Conclusion
The RBT and GAT can offer comparable IOP measures. Therefore, RBT can be recommended for community outreach programs where GAT cannot be employed.


Rwanda J Med Health Sci 2023;6(2):184-190