Validation of the point-of-care (POC) technologies Xpert HIV-1 Qual and m-PIMA HIV 1/2 detect for early diagnosis of HIV-1 and HIV-2 in Senegal


Aissatou Sow Ndoye
Halimatou Diop Ndiaye
Marième Diallo
Aminata Diack
Khadydiatou Coulibaly
Gora Lo
Brianan Kiernan
Pauline Yacine Sène
Ousseynou Ndiaye
Ndèye Fatou Ngom
Mengue Fall
Karim Diop
Sokhna Bousso Gueye
Assane Dieng
Charlotte Lejeune
Cheikh Tidiane Ndour
Coumba Toure Kane

Abstract

Introduction: early infant diagnosis (EID) is crucial in the prevention of mother to child transmission (PMTCT) of human immunodeficiency virus (HIV) and is an essential component for the elimination of HIV. EID can be strengthened in resource-limited countries by the introduction and the roll out of real-time polymerase chain reaction (PCR) technologies via point-of-care (POC) devices which improves treatment in remote areas and reduces turnaround time for clinicians and patients to receive results and linkage to care. The objective of this study was to evaluate the performance of Xpert® HIV-1 Qual Assay (Cepheid) and m-PIMA™ HIV 1/2 Detect (ABBOTT) for EID of HIV-1 and HIV-2.


Methods: the performance of the Xpert® HIV-1 qual device was evaluated with 192 samples including 100 dried blood spot (DBS) samples from the National Reference Laboratory biobank (71 negative and 29 positive samples) and an additional 92 whole blood samples collected from infants from neonatal departments. These infants from seven treatment centers in the Dakar region were born to mothers infected with HIV-1 (n=91), HIV-2 (n= 8) or HIV-1/2 (n=1). The m-PIMA™ HIV 1/2 detect assay was evaluated on whole blood samples (n=100) with 92 HIV-1 samples and 8 HIV-2 samples from children born to HIV-infected mothers. The Cobas AmpliPreP/Cobas TaqMan (CAP/CTM) platform from Roche Diagnostic Laboratories was used as a reference for HIV-1 diagnosis and the Generic HIV-2 Viral Load Assay (Biocentric) was used as a reference for HIV-2 diagnosis. Performance was evaluated by calculating sensitivity (Se), specificity (Sp), positive predictive value (PPV), negative predictive value (NPV) and Cohen's kappa coefficient.


Results: for HIV-1 detection on GeneXpert and m-PIMA, no discordance was found on the samples tested, i.e. a sensitivity of 100% (95% CI: 93.9-100%), a specificity of 100% (95% CI: 97.5-100%), a positive predictive value (PPV) of 100% (95% CI: 93.9-100%) and a negative predictive value (NPV) of 100% (95% CI: 97.5-100%). Agreement with Cobas AmpliPrep/Cobas TaqMan (CAP/CTM) was 100% with a Kappa coefficient of 1 (p<0.001, 95% CI) for both techniques. Similarly, the comparison between m-PIMA and generic biocentric for the detection of HIV-2 on the 8 samples tested showed perfect agreement.


Conclusion: these results confirm the excellent performance of the Xpert® HIV-1 qual and m-PIMA™ HIV1/2 detect tests for the detection of HIV-1 and HIV-2 and encourage the extension of POC tests to improve access to EID in Senegal.