Entrance surface dose of the common radiological investigations in Abuth, Zaria


E Nwokorie
TC Akpa
OP Mallam
E Akpan
PO Ibinaye
OR Obiako
C Nzotta
AC Ugwu

Abstract

The entrance surface Dose (ESD) was determined among patients who presented for routine radiological examinations using thermoluminescent dosimeter (TLD) chips placed on averagely built adult [ (30cm) chest width, 60 - 90kg weight and 20-40years)] by multiplying the incident dose/entrances dose by the back scattered factor [BSF] gotten from an experiment. The values obtained from ESD measurements are as follows 8.03 mGy for Abdomen, 0.13mGy for chest AP, 0.69 mGy for chest lateral, 0.5mGy for cervical spine; AP: 0.38mGy for cervical spine lateral; 4.75mGy for hip AP: 12.83 mGy for hip lateral:5.74mGy for lumbar spine, AP lumbar spine lateral is 15.15mGy. 1.38mGy is for pelvis AP. 2.93mGy for skull AP/PA 1.2mGy for skull lateral, 5.46mGy for thorax AP, while the lateral is 14.90mGy. The average dose value is 9.28mGy Range 0.13-14.90mGy (Percentage standard error; ± 1.45). From this studies radiation doses delivered for routine investigations are generally lower than IAEA reference doses. Using these values for optimization of radiation protection practices for patient should be ensured since it does not rule out options for stochastic radiation.

Keywords: Entrance Surface Dose, (ESD) Routine Radiological Examinations, Reference value, Measured value, Calculated value, Thermolumnient Dosimeter (TLD)