A comparison of sphygmomanometric and oscillometric methods of blood pressure measurements in adults in –patients.


F E Amadusun
J I Isa

Abstract



The Study compared two non-invasive methods of blood pressure measurements used in the anaesthetic management of patients, the conventional sphygmomanometric and the oscillometric methods. One hundred adult in-patients were involved and 400 blood pressure measurements were done with the two devices. The sphygmomanometric measurements were higher in systolic, disastolic and mean arterial pressure values. Correlation coefficients values between the two methods were r= 0.97.0.81.0.95 for systolic, diastolic and mean arterial pressure respectively, (p<0.05).
We conclude that there is a statistically significant difference in haemodynamic values determined by the two devices. This difference is however not clinically important to warrant a recommendation of adjustment when comparing values determined by the two devices.


Keywords: Blood Pressure Measurement, Sphygmomanometer, Oscillometer, Comparison

Annals of Biomedical Science Vol. 3 (1&2) 2004: pp. 61-66